Topology | Symmetric triaxial |
Feedback | Force balance with capacitive transducer |
Mass Centering | Not required |
Self-noise | See graph |
Sensitivity | 750 V•s/m nominal ±0.5% precision |
Off-axis Sensitivity | ±0.5 % |
Bandwidth | -3dB points at 120 s and 100 Hz |
Transfer Function | Lower corner poles within ±0.5% of nominal provided
High-frequency response within 1dB of nominal |
Clip Level | No peak in response at high frequency |
Parasitic Resonances | 26 mm/s from 0.1 Hz to 10 Hz |
Operational Tilt Range | None below 100 Hz
±2.5° without re-leveling |
Technology | Dual degree-of-freedom motorized gimbals
Jam-free mechanism Kinematic design preserves full seismometer |
Clamping | performance |
Range | Active, sequenced by gimbals processor |
Accuracy | >±180° relative to upright case |
Leveling Initiation | Levels to within ±0.5° of true vertical
Some or all of: • Configurable delay after power on • Configurable periodic • Automatic as-required, based on mass positions • External Command |
Type | RS-232 compatible Serial Line Internet Protocol (SLIP) |
Leveling | On board Web Server, standard HTTP
Initiate immediate leveling via user interface Automatic leveling mode (post power-on, periodic, as-needed) |
On Request | Independent mass position values
Instrument temperature SOH Magnetometer readings Case orientation (with respect to vertical) Seismometer orientation (with respect to vertical) Download state-of-health (SOH) log Erase SOH log Download custom information Factory sensitivity and sensor response data User calibration data (poles and zeros) Instrument serial number and firmware revision |
Leveling Log | Every leveling event is logged in non-volatile memory
Full before-and-after state-of-health is logged |
State-of-Health Log | Scheduled interval logging of SOH, including:
• Time from power-on • Seismometer mass positions • Vessel and seismometer orientations • Magnetometer readings • Temperature Internal relative humidity Capacity for more >2 years of twice daily level checks and SOH logging |
Enclosure | Stainless Steel cylinder
Connector on top end cap |
Diameter | 151 mm |
Height | 258 mm, not including connectors |
Weight | 14.5 kg on land |
Main Connector | 12-pin female, VSK-12-BCL rubber-molded glass epoxy
40V peak-to-peak differential seismic signal plus ground (3 channels) Serial RS-232 port (Rx, Tx, GND) Calibration voltage input Power input and return |
Supply Voltage | 9 to 29 VDC isolated |
Power Consumption | <180 mW typical (leveled, quiescent) |
Protection | Reverse-voltage and over-voltage protected
Self-resetting over-current protection Unit can be powered on during installation and retrieval |
Operating Temperature | -20ºC to +60ºC |
Storage Temperature | -40ºC to +70ºC |
Shock | 100g half sine, 5 ms without damage, 6 axes |
Transportation | No seismometer mass lock required prior to deployment and through full experiment cycle |
Magnetic | Insensitive to natural variations of the Earth’s magnetic field |
Specifications subject to change without notice